Model Number 353/19004
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Dyspnea (1816)
Event Date 01/09/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the device was not having any visible damage, but the patients complained about breathlessness.There were no complications as they got connected to a new device immediately.
 
Manufacturer Narrative
Evaluation summary: the reported device used was not retained by the customer and therefor not available for evaluation.Eleven boxes containing two-hun dred-seventy-five unused samples still in their original sealed packaging were received.A flow resistance test was performed on the samples to determine pressure resistance in the filters.All samples were found to be within specification limits.The investigation found all devices functioned as intended.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAR
Type of Device
CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)
Manufacturer (Section D)
MALLINCKRODT DAR SRL
via giacomo bove 2/4/6/8
mirandola 41037
IT  41037
MDR Report Key9616538
MDR Text Key175899784
Report Number2936999-2020-00057
Device Sequence Number1
Product Code BYD
Combination Product (y/n)N
PMA/PMN Number
K941535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number353/19004
Device Catalogue Number353/19004
Device Lot Number19K1050FAX
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/16/2020
Initial Date Manufacturer Received 01/09/2020
Initial Date FDA Received01/22/2020
Supplement Dates Manufacturer Received02/24/2020
Supplement Dates FDA Received03/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-