Model Number 2217-50-041
Device Problem Material Twisted/Bent (2981)
Patient Problem Not Applicable (3189)
Event Date 01/08/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the straight cup impactor cross threaded onto the pinnacle cup.The cup was removed on the back table and a new handle was opened and the case was completed.There was a 5 minutes surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. .
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary : examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9616958
MDR Text Key189247821
Report Number1818910-2020-02513
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295098980
UDI-Public10603295098980
Combination Product (y/n)N
PMA/PMN Number
P070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2217-50-041
Device Catalogue Number221750041
Device Lot NumberSO2037982
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Initial Date Manufacturer Received 01/08/2020
Initial Date FDA Received01/22/2020
Supplement Dates Manufacturer Received01/30/2020
02/19/2020
Supplement Dates FDA Received02/13/2020
02/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age50 YR
-
-