The reported tear/rip/hole was confirmed during returned device analysis.A review of the lot history record identified no exception issued to the reported lot.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on the information reviewed and the return device analysis, the reported tear/rip/hole was confirmed due to the observed tear in the packaging.Therefore, this appears to be a potential product issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
|