Analysis found that there was a residue consistent with biological contaminants on both devices.The inner assembly of both samples would turn freely by hand however sample 1 inner tip broke off at the spiral wrap.The spiral wrap was twisted in on itself in a clockwise direction 0.79¿ from tip indicating torsion load and aggressive use.The tips of both samples were compacted with biological material with varying degrees which may require the user to apply more pressure to maintain performance.The distal end of the outer tubes inside diameter of both samples was rough and worn, sample 1 wore through the outer tube wall.The adjacent inner assemblies were rough and worn on both samples.If information is provided in the future, a supplemental report will be issued.
|