LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/27/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed two error codes associated to the patient side during service.There was no report of patient injury.
 
Manufacturer Narrative
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported error codes.He traced the fault back to both patient pumps.He replaced the pumps and the issue was solved.Functional verification testing was completed without further issues and the unit was returned to service.The replaced parts were investigated and rust and oxidation on the motor bearing as well as visible residue on impellers and on internal components could be identified.The root cause of the reported event has been traced back to rust and oxidation on the bearing mechanisms as well as residue on impellers and on internal components, due to environmental condition, not allowing the pump motors to rotate freely.Through a device service history analysis under another event involving the same device and reporting the same failure, livanova found out that a duplicate was submitted.The case received and already reported under medwatch # mw-007857 has been voided and this has been taken to capture all the above mentioned information.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9646528
MDR Text Key196432130
Report Number9611109-2020-00038
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/31/2019
Initial Date FDA Received01/30/2020
Supplement Dates Manufacturer Received02/26/2020
Supplement Dates FDA Received03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-