• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problems Display or Visual Feedback Problem (1184); Smoking (1585)
Patient Problem No Patient Involvement (2645)
Event Date 01/02/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.(b)(4).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.He traced the failure back to the inverter board of the touchscreen, in particular to the coil.Since this type of board is obsolete, the technician replaced the touchscreen and the associated parts.The issue was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
 
Event Description
Livanova (b)(4) received a report that during setup a s5 system alarmed.The s5 double head pump was restarted and smoke came out of the device.The pump was swapped with another.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9646573
MDR Text Key176949742
Report Number9611109-2020-00041
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/02/2020
Initial Date FDA Received01/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2008
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-