SORIN GROUP ITALIA S.R.L. - CRM FACILITY REPLY; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number REPLY DR |
Device Problem
Over-Sensing (1438)
|
Patient Problems
Fainting (1847); Dizziness (2194)
|
Event Date 09/27/2019 |
Event Type
Injury
|
Event Description
|
Reportedly, the patient, who is pacemaker-dependent, complained about fainting and dizziness since the follow-up performed on (b)(6) 2019.Review of patient files retrieved during the follow-ups performed on (b)(6) 2019, (b)(6) 2019, (b)(6) 2019 and (b)(6) 2019 revealed noise on both atrial and ventricular channels.No provocative maneuvers were performed during the follow-up dated (b)(6) 2019, as no doctor was present.Besides, the programmed settings remained unchanged: pacing mode set in safer with a basic rate at 60 min-1, atrial pacing output at 2.0 v/ 0.35ms and ventricular pacing output at 2.5 v/ 0.35 ms.During the follow-up performed on (b)(6) 2019, the pacing mode was re-programmed to doo.During the follow-up dated (b)(6) 2019, the ventricular pacing output was re-programmed to 4.0 v/0.35 ms.The next follow-up was performed on (b)(6) 2020.Preliminary analysis confirmed simultaneous noise oversensing on both channels.The origin of these non-physiological signals is unknown.It could be located at lead(s) level, connection(s) level or pacemaker level; none of them can be confirmed with available data.
|
|
Event Description
|
Reportedly, the patient, who is pacemaker-dependent, complained about fainting and dizziness since the follow-up performed on (b)(6)2019.Review of patient files retrieved during the follow-ups performed on (b)(6)2019 27 (b)(6)2019 ,(b)(6)2019 and (b)(6)2019 revealed noise on both atrial and ventricular channels.No provocative maneuvers were performed during the follow-up dated (b)(6)2019 , as no doctor was present.Besides, the programmed settings remained unchanged: pacing mode set in safer with a basic rate at 60 min-1, atrial pacing output at 2.0 v/ 0.35ms and ventricular pacing output at 2.5 v/ 0.35 ms.During the follow-up performed on 27 (b)(6)2019 , the pacing mode was re-programmed to doo.During the follow-up dated (b)(6)2019 , the ventricular pacing output was re-programmed to 4.0 v/0.35 ms.The next follow-up was performed on 23 january 2020.Preliminary analysis confirmed simultaneous noise oversensing on both channels.The origin of these non-physiological signals is unknown.It could be located at lead(s) level, connection(s) level or pacemaker level; none of them can be confirmed with available data.
|
|
Manufacturer Narrative
|
Please refer to the attached analysis report.
|
|
Search Alerts/Recalls
|
|
|