DEPUY MITEK LLC US FOOT PEDAL(FMS VUE/NEXTRA); FOOT-SWITCH, ELECTRICAL Back to Search Results
Model Number 283573
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/24/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported via the affiliate that the fms was pulled out to use in an a/s case.Blood and saline was leaking out of the back of the footpedal.Level 2 ortho nurse said that she had asked ns whether they had been bagging the pedal for cases and they said they had.Ns reported that there was a considerable amount of bloody and fluid leaking out.Footpedal discarded as bmt would not look at footpedal due to amount of blood, and cssd also could not decontaminate due to electrics.Hospital used an alternative footpedal to complete cases.No delay was reported.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that during that the blood and saline was leaking out of the back of the footpedal, that there was a considerable amount of bloody and fluid leaking out.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number: g38ay1772, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: a manufacturing record evaluation was performed for the finished device lot number: g38ay1772, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOT PEDAL(FMS VUE/NEXTRA)
Type of Device
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9665085
MDR Text Key204070625
Report Number1221934-2020-00466
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705017010
UDI-Public10886705017010
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283573
Device Catalogue Number283573
Device Lot NumberG38AY1772
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/28/2020
Initial Date FDA Received02/04/2020
Supplement Dates Manufacturer Received04/17/2020
Supplement Dates FDA Received04/20/2020
Patient Sequence Number1
-
-