It was reported that the patient had high lead impedance.However, 2 further diagnostics were performed that day and the impedance returned ok.Then, approximately 1 month later, the patient was found to have high impedance again.No adverse events were reported related to the high impedance.No known surgical intervention has occurred to date.No additional relevant information has occurred to date.
|
Clinic notes were received indicating that shortly after generator replacement, at the time that high impedance was identified, there was a lack of efficacy with the newly implanted device.The lack of efficacy can be attributed to the temporary interruption in therapy, as the patient's previous generator battery had depleted, and the patient's settings had to be re-titrated to the prior therapeutic settings.As the settings were increased, the patient's depression improved.However, it is also possible that the efficacy was adversely impacted by the high impedance.No additional relevant information has been received to date.No known surgical intervention has occurred to date.
|