At 09:11 hours, ipp9 sent an error message to the monitors of: "limited table/stand movement".The table and ap flat detector moved significantly slow like it was in collision mode.All of the resets were complete with troubleshooting the room and the error did not clear.At this time, clinical imaging was in the core and witnessed with issue.A hard shut down and reboot of the room was then completed.The error still didn't clear, and the physician could not proceed with the procedure in ipp9.The intubated patient was then moved to another room for completion of the procedure.A service call was placed to siemen's and the service clinician came to assess the room.Manufacturer response for system, x-ray, fluoroscopic, image-intensified, (artis vd 11c) (per site reporter).The rep from siemens has not arrived yet.
|