Model Number ACU0T0
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 11/27/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A health professional reported that during an intraocular lens (iol) implant surgery, the tip of the lens was deformed.The surgeon did not implant the lens.Additional information has been requested.
 
Manufacturer Narrative
Additional information provided in d.10., h.3., h.6., and h.10.The device with the lens was returned.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger is at the trailing optic edge.The lens is advanced to mid-nozzle.The trailing haptic is folded in on the optic.The leading haptic is extended.All product and batch history records are quality reviewed prior to product release.A non-qualified viscoelastic was indicated.No problems are observed with the returned device.The plunger, lens and haptic positions are acceptable.The leading haptic is straight.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the dfu.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9676190
MDR Text Key182363619
Report Number1119421-2020-00190
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model NumberACU0T0
Device Lot Number12594574
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Initial Date Manufacturer Received 01/13/2020
Initial Date FDA Received02/06/2020
Supplement Dates Manufacturer Received03/26/2020
Supplement Dates FDA Received03/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
HEALON
-
-