The reporter indicated that a vticm5_13.7, -7.50/2.5/083 (sphere/cylinder/axis), implantable collamer lens was implanted into the patients left eye (os) on (b)(6) 2019.Excessive vault, significant reduction of irido-corneal angles and pigment dispersion were reported.The reporter also stated that "iris has been damaged at the limbal region, where you can see that the pigment layer has been dissolved." secondary surgical intervention (yag) was performed on post-op day 1 "because the condition of the eye and patient's rapport (headache and vomiting debuting ca 8 h after surgery) suggested that iop had been high." in the surgeons opinion the cause of the event was due to icl too large/ actual sulcus diameter smaller than expected from preop measurements.For patients' current condition the surgeon reported " patient is very pleased with the outcome, wears no lenses or glasses and would like to keep everything as it currently is; so close to emmetropia oa.However, there is rather large amounts of pigmentation on icl, iris atrophy and the vault is still ca 4 oa".The surgeon plans to remove the icl with no icl implantation." if additional information is received a supplemental medwatch report will be submitted.
|
H6- device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order (s), including the suspected device, have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contribute to the complaint issue.Claim # (b)(4).
|