• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) VITAL SYNC REMOTE MONITORING SYSTEM; TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) VITAL SYNC REMOTE MONITORING SYSTEM; TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY Back to Search Results
Model Number 10116080
Device Problems Device Alarm System (1012); Computer Software Problem (1112); Filtration Problem (2941)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 01/24/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the unit did not sound an alarm for "filterline disconnected" or "sensor disconnected" upon initialization.There was no allegation of patient death or serious injury associated with this event.
 
Manufacturer Narrative
Additional information: b5, d3(mfr name, street 1, mfr city, country code, postal code), g4, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the software was tested to research the mirroring parameters.The device would not sound an alarm for "filter line disconnected" or "sensor disconnected" while it does on software so it was changed to mirror the device setting that both of them should not alarm.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITAL SYNC REMOTE MONITORING SYSTEM
Type of Device
TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez MN 32590
MX  32590
MDR Report Key9683717
MDR Text Key182768101
Report Number2936999-2020-00103
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
K093422
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 05/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10116080
Device Catalogue Number10116080
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/25/2020
Initial Date FDA Received02/07/2020
Supplement Dates Manufacturer Received05/06/2020
Supplement Dates FDA Received05/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-