• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS) Back to Search Results
Model Number 37612
Device Problems High impedance (1291); Migration or Expulsion of Device (1395); Pocket Stimulation (1463); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Tingling (2171); Electric Shock (2554)
Event Date 01/27/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the ins was no longer fixed in the pocket and was able to be moved up and down approximately 2 cm.When it was moved down, they felt a slight tingling/burning sensation.Diagnostics performed were turning stimulation off to confirm that the tingling was dependent on stimulation.An impedance check was also performed in various ins positions, however they did not show any variations.The patient decided to continue to use the device as usual as the sensation was acceptable for them.The issue was considered resolved at the time of the report.It was also noted there were high, out of range impedances on the patient's right side for contact 11.
 
Manufacturer Narrative
B3: event date was initially reported as (b)(6) 2020; however, additional information indicated the event started earlier than previously reported.Year is known, but exact month and date are not known.H6: method code 4117 has replaced 4114.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received.It was noted the patient had been experiencing the shocking/jolting sensation for a couple of weeks and not suddenly.The sensation only appears when the ins is moved down, but there were no known falls or accidents associated with the event.The physician had not reprogrammed the patient, although this had been suggested to them.The contact with high impedance was not used for therapy.The cause of high impedance and shocking sensation was not known; previously,impedance was normal.No actions had been taken, and the patient still had adequate therapy.
 
Event Description
It was clarified that the shocking sensation was in fact a burning/tingling sensation.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777 1200
MDR Report Key9688338
MDR Text Key188796776
Report Number3007566237-2020-00215
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2015
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/06/2020
Initial Date FDA Received02/10/2020
Supplement Dates Manufacturer Received02/10/2020
02/18/2020
10/13/2020
Supplement Dates FDA Received02/18/2020
03/03/2020
10/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age23 YR
-
-