• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problems Overheating of Device (1437); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/08/2020
Event Type  malfunction  
Manufacturer Narrative
A product sample was received and it is awaiting evaluation.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during an ocular procedure, the surgeon indicated he felt that the laser was hotter or treatment was too effective at times.The procedure was completed with the same laser probe and system without harm to the patient.It was also reported that the main foot pedal was sticky.Additional information was requested; however, the reporter indicated that there is no further information available.
 
Manufacturer Narrative
Additional information has been provided in d.10, h.3, h.6, and h.10.The company service representative examined the system and was unable to duplicate the problem reported.The footswitch was replaced as a prevention.The system was then tested and met all product specifications.The footswitch was received and a visual assessment did not find any non-conformities.Further evaluation found no problem with the footswitch.Based on assessment, the product met specifications at the time of release.The root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key9689770
MDR Text Key182780708
Report Number2028159-2020-00078
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2020
Initial Date Manufacturer Received 01/15/2020
Initial Date FDA Received02/10/2020
Supplement Dates Manufacturer Received04/22/2020
Supplement Dates FDA Received05/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
LASER PROBE
-
-