BAXTER HEALTHCARE CORPORATION 500 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938738
Device Problems Fluid/Blood Leak (1250); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The lot was manufactured from march 28, 2019 - march 29, 2019.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 500ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was leaking from unspecified location.The amino acid solution was noted to be discolored.This was discovered during mixing.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was received that the statement ¿the amino acid solution was noted to be discolored¿ was erroneously added.This statement has been removed.The device was received for evaluation.Unaided visual inspection was performed which did not identify any abnormalities that could have contributed to the reported condition.A functional testing was performed which revealed a leak at the spike port cap from the spike port.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
500 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key9696206
MDR Text Key178740681
Report Number1416980-2020-00599
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938738
Device Lot Number60177107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2020
Initial Date Manufacturer Received 01/15/2020
Initial Date FDA Received02/11/2020
Supplement Dates Manufacturer Received02/19/2020
Supplement Dates FDA Received03/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-