The reported event was confirmed as manufacturing related.Visual evaluation of the returned sample noted one opened (without original packaging), 3-way temperature sensing silicone foley catheter.Visual inspection of the sample noted a large piece of flash (measuring 0.1640 inches) on the shaft located 3.7835 inches from the tip of the catheter.The device history record was reviewed and found a possible manufacturing issue(s) that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.".
|