• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGTRS60AMT
Device Problems Entrapment of Device (1212); Separation Failure (2547); Failure to Form Staple (2579)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/10/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during second fire resection on the stomach on a laparoscopic sleeve gastrectomy, the stapler was able to fire however there were incomplete staple line proximally.In addition, the device jaw locked on tissue but was opened normally.The surgeon had to retire the reload to cut manually the reinforcement and use another device to complete the case.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one devices.Visual inspection of the returned product noted that the reload was partially fired with the interlock engaged.The jaws were open.Staple pushers were visible at the 4cm cut line.Microscopic evaluation noted that the reload had damage to the cutting edge of the knife blade.The trs material was properly anchored to the anvil and cartridge at the distal end of the cartridge.The distal sutures were intact.Functionally, the reload was loaded into a post market vigilance instrument, the interlock was overridden, and the reload was cycled without hesitation and was applied to test media.All remaining staples were placed, and test media was transected.The trs material was properly placed and sutures were properly cut.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the damaged knife blade may occur when an obstacle has been incorporated in the jaws during application.The information booklet which accompanies each product shipment cautions the user; ensure that no obstructions (such as clips) are incorporated in the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key9699430
MDR Text Key179127320
Report Number1219930-2020-00673
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884521543638
UDI-Public10884521543638
Combination Product (y/n)N
PMA/PMN Number
K160176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberSIGTRS60AMT
Device Catalogue NumberSIGTRS60AMT
Device Lot NumberN8G0387RX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2020
Initial Date Manufacturer Received 01/29/2020
Initial Date FDA Received02/12/2020
Supplement Dates Manufacturer Received03/18/2020
Supplement Dates FDA Received04/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age30 YR
Patient Weight140
-
-