• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH METASUL DUROM, COMPONENT FOR ACETABULUM, UNCEMENTED, 56/A¸ 50, CODE P; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH METASUL DUROM, COMPONENT FOR ACETABULUM, UNCEMENTED, 56/A¸ 50, CODE P; N/A Back to Search Results
Model Number N/A
Device Problem Material Fragmentation (1261)
Patient Problems Failure of Implant (1924); Test Result (2695)
Event Date 12/17/2019
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products and therapy date detail of product: item # 0100181500, item name metasul ldh, head,50, code p, taper 18/20, lot # 2481361.Item # unknown, item name head adapter hip, lot # unknown.Item # unknown, item name alloclassic stem hip, lot # unknown.The manufacturer did not receive devices or x-rays for review.As no lot numbers were provided for the devices, the device history records could not be reviewed.Based on an extensive investigation of events reported from several user facilities outside the usa, zimmer identified that the most probable cause for the outcome observed was a loose or unstable cup that resulted from use of surgical techniques not consistent with the manufacturer's recommendations.As a corrective action, a retraining program for users outside the usa was initiated in november 2009 and reported to the national competent authorities as required.The durom cup reported in this case is not marketed in the usa.A similar cup, compatible with the metasul ldh femoral head, is cleared in the usa and a corrective action for this product was reported to the fda in july 2008 as notification z-2415/2426-2008.No further investigation required as this issue is known and addressed in (b)(4) (error pattern: potential early revision of the acetabular component due to loosening, implant migration or unresolved pain, higher ion release).At least one of these error patterns is observed in this event.Should any additional information that changes the assessment become available to us, or any extra demand be requested, we will re-evaluate the case.The need for corrective measures is not indicated and zimmer (b)(4) considers this case as closed.(b)(4).
 
Event Description
It was reported that patient underwent revision surgery due to metallosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METASUL DUROM, COMPONENT FOR ACETABULUM, UNCEMENTED, 56/A¸ 50, CODE P
Type of Device
N/A
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key9701625
MDR Text Key191152110
Report Number0009613350-2020-00063
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number01.00214.056
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2019
Initial Date FDA Received02/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-