• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Catalog Number AK-05502
Device Problem Shipping Damage or Problem (1570)
Patient Problem No Patient Involvement (2645)
Event Date 02/03/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that there have been around 5 kits that have shown up recently either with damage to the outer tray or the lid stock partially torn off, making product not sterile for use.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the outer tray and kit with no relevant findings.The customer reported the outer tray of the kit was damaged.The customer returned damaged kit.Visual examination of the returned kit revealed the lower left and right corners of the outer tray are stressed and the lower left corner was cracked under the flange.No other defects or anomalies were observed.The customer also provided photos that show the corner of kit having a crack.A corrective action is not required at this time as the investigation shows no evidence to suggest a manufacturing related cause.Per the manufacturing site, all incoming outer trays are visually inspected for evidence of stress marks prior to packaging as well as each packaged kit is inspected after sealing and prior to being placed into the corrugate container.The reported complaint that a tray was cracked was confirmed based on the sample received.Visual inspection of the returned kit revealed the lower left and right corner showed signs of stress and the left corner had a crack under the flange.The customer also provided photos that show the corner of kit having a crack.A device history record review was performed on the outer tray and kit with no relevant findings.Also, the outer trays are 100% inspected prior to and after packaging.It is unknown how the kit was handled during shipping or prior to use.The investigation found no evidence to suggest a manufacturing related cause.Therefore, the potential cause of the tray's corners being stressed , and cracking could not be determined.
 
Event Description
It was reported that there have been around 5 kits that have shown up recently either with damage to the outer tray or the lid stock partially torn off, making product not sterile for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9702068
MDR Text Key179284363
Report Number1036844-2020-00072
Device Sequence Number1
Product Code BSO
Combination Product (y/n)N
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Catalogue NumberAK-05502
Device Lot Number23F19J0104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2020
Initial Date Manufacturer Received 02/03/2020
Initial Date FDA Received02/12/2020
Supplement Dates Manufacturer Received03/20/2020
Supplement Dates FDA Received03/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-