• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEATER COOLER UNIT; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEATER COOLER UNIT; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701034642 - HCU 30
Device Problems Circuit Failure (1089); Electrical /Electronic Property Problem (1198); Electrical Shorting (2926)
Patient Problem No Patient Involvement (2645)
Event Date 02/10/2020
Event Type  malfunction  
Manufacturer Narrative
When further information becomes available a follow up emdr will be submitted.
 
Event Description
It was reported that the hcu30 causing that the breaker tripped.The heater had an eletrical leak.Complaint # (b)(4).
 
Manufacturer Narrative
It was reported that the hcu 30 causing the breaker in the operation room tripped.According to the field service technician on 2020- 02-18 via communication grid the hcu 30 is at the local getinge office.The technician found, that the 701034044 mcp00704749 # heater card 230v had a short circuit.According to the service report#43285926 dated on 2020-03-16 the maquet service technician replaced the 701034044 mcp00704749 # heater card 230v.All function tests passed.The device is back in clinical use.The most probable root cause of the tripped breaker in the operation room is a short circuit on the hcu30 heater card 230v (material#70103.4044).Thus the reported failure "breaker tripped" can be confirmed.The reported failure happened during start up of the device.The hcu 30 which was used, was responsible for this complaint.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrance rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER UNIT
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key9715909
MDR Text Key200421606
Report Number8010762-2020-00064
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K031554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 03/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701034642 - HCU 30
Device Catalogue Number701034642
Initial Date Manufacturer Received 02/10/2020
Initial Date FDA Received02/17/2020
Supplement Dates Manufacturer Received03/25/2020
Supplement Dates FDA Received03/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-