Model Number 1050044
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2020
Event Type  malfunction  
Manufacturer Narrative
Pma/510k: reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K111959.If additional information becomes available a follow up report will be submitted.
 
Event Description
It was reported that the ampoule leakage occurred.The reporter indicated when opening the package, the leakage from the ampule was already caused, therefore the product was not able to be used.
 
Manufacturer Narrative
Investigation: samples received: (b)(4) open pouch, ampoule unopened with leakage signs.Analysis and results: there are previous complaints of this code batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical warehouse.We have received one open pouch (closed ampoule) showing ampoule leakage.The ampoule received has been optically evaluated and a defect in the sealing bar of the ampoule was found.The leakage of the glue occurs at this point.Reviewed the batch manufacturing record, this product had an incidence not related to this issue and was released fulfilling b.Braun surgical specifications.Final conclusion: taking into account that the results of the sample received do not fulfil b.Braun surgical specifications, and that there are several previous complaints for the same defect, we conclude that the complaint is confirmed by evidence of failure in the sample received.Actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, we have opened a capa.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HISTOACRYL BLUE 0.5ML
Type of Device
TISSUE ADHESIVES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key9717668
MDR Text Key179727737
Report Number3003639970-2020-00097
Device Sequence Number1
Product Code MPN
Combination Product (y/n)N
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1050044
Device Catalogue Number1050044
Device Lot Number219031N1
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/29/2020
Initial Date FDA Received02/17/2020
Supplement Dates Manufacturer Received03/26/2020
Supplement Dates FDA Received04/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-