Model Number 37800
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/07/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) regarding a patient with an implantable neurostimulator (ins).It was reported that they ran impedances and found the values to be high.The hcp stated they programmed the patient using c+ 2- to avoid the high impedances and sent the patient for x-rays.The following values were provided: 2c 3703c >2000023 20000.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the hcp.It was reported that the device was surgically replaced/revised.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
Product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2014, product type: lead; product id: 4351-35, serial# (b)(4), implanted: (b)(6) 2014, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from a healthcare provider (hcp).It was reported that the lead was surgically replaced and the patient is doing great.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key9728538
MDR Text Key180124609
Report Number3004209178-2020-03656
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/07/2020
Initial Date FDA Received02/19/2020
Supplement Dates Manufacturer Received03/03/2020
04/15/2020
Supplement Dates FDA Received03/05/2020
04/17/2020
Date Device Manufactured08/31/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age35 YR
-
-