Model Number 8900 |
Device Problems
Communication or Transmission Problem (2896); Detachment of Device or Device Component (2907)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 02/07/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.
|
|
Event Description
|
It was reported that a tip detachment occurred.A percutaneous coronary intervention was being performed.Vascular access was obtained via the radial artery.The target lesion was located in the mid right coronary artery (rca).This comet pressure guidewire along with an entry needle were advanced to the target lesion.However, a tip detachment occurred.The detached tip was unable to be retrieved and was secured in the patient coronary vessel.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.The devices shaft was visually and microscopically inspected for damage.The device showed 2 small bends located 16cm and 22cm from the separated distal end.The device showed a separation located 3.2cm from the tip.The total length of a comet wire is 185cm.The returned segment measured 181.8cm which means that 3.2cm was not returned.Functionality of the device could not be confirmed due to the damage of the device.The coefficient was confirmed to be programmed per specification.A material testing analysis and characterization was performed and identified that the proximal end showed that a fracture occurred at the beam location.Striations were present on the fractured surface which suggest reverse bending fatigue with possible ductile overload.Mechanical damage was also present at the final ruptured location and on some of the fractured surface.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that a tip detachment occurred.A percutaneous coronary intervention was being performed.Vascular access was obtained via the radial artery.The target lesion was located in the mid right coronary artery (rca).This comet pressure guidewire along with an entry needle were advanced to the target lesion.However, a tip detachment occurred.The detached tip was unable to be retrieved and was secured in the patient coronary vessel.The procedure was completed with another of the same device.No patient complications were reported.It was further reported that the lesion was 95% stenosed and located in the tortuous and calcified, 25mm in length radial artery.The separation occurred after multiple attempts were made to cross the lesion.The break occurred 3cm from the tip and the tip became stuck in the lesion.At the time of the separation the guidewire had not prolapsed and had maintained position in the ostium during the entire procedure; however, the pressure signal was lost.
|
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.
|
|
Event Description
|
It was reported that a tip detachment occurred.A percutaneous coronary intervention was being performed.Vascular access was obtained via the radial artery.The target lesion was located in the mid right coronary artery (rca).This comet pressure guidewire along with an entry needle were advanced to the target lesion.However, a tip detachment occurred.The detached tip was unable to be retrieved and was secured in the patient coronary vessel.The procedure was completed with another of the same device.No patient complications were reported.It was further reported that the lesion was 95% stenosed and located in the tortuous and calcified, 25mm in length radial artery.The separation occurred after multiple attempts were made to cross the lesion.The break occurred 3cm from the tip and the tip became stuck in the lesion.At the time of the separation the guidewire had not prolapsed and had maintained position in the ostium during the entire procedure; however, the pressure signal was lost.
|
|
Search Alerts/Recalls
|