MAUDE Adverse Event Report: BIOMET UK LTD. DELTA CER FEM HD 28/+3MM T1; CERAMIC FEMORAL HEAD PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. DELTA CER FEM HD 28/+3MM T1; CERAMIC FEMORAL HEAD PROSTHESIS Back to Search Results
Model Number N/A
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a supplemental mdr will be submitted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the size of biolox delta head indicated on the implant (28mm ceramic head -3) is different compare to the size indicated on the package (28mm ceramic head +3).Patient was being operated on but not affected.5 minute delay of surgery, as surgeon was checking the implant.
 
Event Description
It was reported that the size of biolox delta head indicated on the implant (28mm ceramic head -3) is different compare to the size indicated on the package (28mm ceramic head +3).Patient was being operated on but not affected.5 minute delay of surgery, as surgeon was checking the implant.
 
Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.As the product has not been received, the investigation was limited to the information provided; a review of device history records and complaint history.The reported event has been determined to be a coob.The event reports that the labelling on the outer box did not match the device.This was identified during surgery and resulted in a 0-15 min extension to surgery time.An alternative device was used to complete the surgery.No photographs of the implant (photographs of the label have been provided) have been provided and no product has been returned, however, visual inspection of the photographs provided through another complaint ((b)(4)) confirm a commingle has occurred.(b)(4) has been raised to hold product affected by the issue.Issue evaluation (b)(4) has been raised to further investigate the issue.Health hazard evaluation (determination) (b)(4) was raised which was escalated to (b)(4).A review of the complaint database over the last 3 years has found no similar complaints reported with these items.A review of the manufacturing history records confirms no abnormalities or deviations reported.Without the opportunity to examine the complaint product, root cause cannot be determined due to insufficient information.If any additional information becomes available, then the complaint will be reopened and investigated thoroughly.H3 other text : product not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA CER FEM HD 28/+3MM T1
Type of Device
CERAMIC FEMORAL HEAD PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
MDR Report Key9743629
MDR Text Key198544629
Report Number3002806535-2020-00093
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
PMA/PMN Number
K131684
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number650-1157
Device Lot Number2019091286
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/29/2020
Initial Date FDA Received02/24/2020
Supplement Dates Manufacturer Received04/14/2020
Supplement Dates FDA Received04/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-