Model Number SCORPION-MULTIFIRE NEEDLE |
Device Problems
Break (1069); Difficult to Remove (1528)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 02/13/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
|
Event Description
|
It was reported by the sales rep, that during a rotator cuff repair procedure, upon insertion the tip of the ar-13995n, needle broke off.The broken piece remains in the patient.
|
|
Manufacturer Narrative
|
The failure mode could not be determined, but the broken needle point condition can be caused by consistently passing the needle through challenging tissue (thick or calcified) or hitting bone.The buckled needle strip condition is typically caused by the device skiving under thick tissue or distorting distally under the jaw.The distal end of the nitinol point demonstrated minimal scrape marks, indicating the device was not fired excessively.The mating (instrument) was not returned or identified.Confirmed the needle strip thickness and width dimensions to be within specification.
|
|