• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT Back to Search Results
Model Number 809810
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/05/2020
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's technical support specialist, the unit is at its end of service life (eosl).Per the user facility's biomedical technician, the unit has been removed from service until a replacement can be installed.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit read over 1000 on the current leakage test, causing the user facility's circuit breaker to trip and the isolation monitor light emitting diode (led) to flash.Also, the screen on the unit went blank.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Updated blocks: h3 and h6.The reported complaint is not verifiable.Multiple diligence attempts have been unsuccessful for part return so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9759681
MDR Text Key200801423
Report Number1828100-2020-00088
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K071521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number809810
Device Catalogue Number809810
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/05/2020
Initial Date FDA Received02/26/2020
Supplement Dates Manufacturer Received03/16/2020
Supplement Dates FDA Received03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-