Catalog Number 4710500394-3 |
Device Problem
Improper Chemical Reaction (2952)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source, foreign - event occurred in (b)(6).The review of the device manufacturing quality record indicates that (b)(4) products optipac 40 refobacin bone cement r-3, reference 4710500394-3, batch 906ba03500 were manufactured on 03rd september 2019.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.No non conformity or deviation was observed which could be linked to the event described in the complaint.The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
|
|
Event Description
|
It was reported that the cement became hard and rubbery in consistency too quickly.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.The device was not returned to the manufacturer.Therefore it could not be analyzed.Reserve sample from the same lot was tested under standardized conditions.The product did not show any unusual behavior during mixing, handling or setting.The reported behavior of the cement could not be reproduced.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.Within 3 years, no similar complaint has been recorded regarding a cement paste too short setting time on optipac 40 refobacin bone cement, reference 4710500394-3, batch 906ba03500.According to available data, root cause of the event was unable to be determined.However, there is no evidence that the event is related to the product.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the cement became hard and rubbery in consistency too quickly.No adverse events have been reported as a result of the malfunction.
|
|