MAUDE Adverse Event Report: COOK INC COOK SPECTRUM; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK SPECTRUM; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Catalog Number C-UTLM-701J-RSC-ABRM-HC-RD
Device Problem Partial Blockage (1065)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2020
Event Type  malfunction  
Manufacturer Narrative
Brand name: cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
It was reported that a cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter was found to be occluded during infusion.The device was being used in the left femoral vein.It was reported "the physician felt resistance during the procedure".All lumens were reported to be filled with saline before introduction into the body.Nutrition medications were being delivered through the device.Saline was used to flush the unused lumen and lock the device.The patient previously had brain surgery.No other adverse effects have been reported for this incident.
 
Event Description
No additional patient/event details have been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable.D10 product received on: 02mar2020.Investigation evaluation.It was reported that a cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set (c-utlm-701j-rsc-abrm-hc-rd) from lot 9510505 became occluded seven days after placement.All lumens were filled with saline prior to introduction into the body.Nutrition medicine was being delivered through the device, and unused lumens were flushed and locked with saline.Cook became aware of this event on (b)(6) 2020 upon being notified by a doctor from national cheng kung university hospital.The patient reportedly experienced no adverse effects as a result of this incident.A review of the complaint history and device history record, as well as a functional test of the returned device were conducted during the investigation.The complaint device was returned with the distal tip cut off.A cook microclave was attached to the white hub, while two third party microclaves were attached to the red and blue hubs.The device examination found that all three hubs were able to be flushed without issue, and biological matter was flushed out of the blue lumen.Additionally, a document based investigation evaluation was performed.A review of the device master record found that sufficient controls are in place to detect this failure mode prior to release.A review of the design history file found that the risks associated with the devices are acceptable when weighed against the benefits.A review of the device history record found no nonconformances or additional complaints associated with this device lot.There is no evidence to suggest there is any nonconforming product in house or out in the field.A review of the product labeling for the device was completed.The instructions for use state the following instructions related to the reported failure mode: suggested catheter maintenance to prevent clotting or possibility of air embolus, the double-lumen¿s #2 lumen, the triple-lumen¿s #2 and #2 lumens, and the five-lumen¿s #2, #3, #4, and #5 lumens should be filled with saline solution or heparinized saline solution (100 units of heparin per ml of saline is usually adequate), depending on institutional protocol, prior to catheter introduction.Any unused lumens should be maintained with continuous saline or heparinized saline drip or locked with heparinized saline solution.Before using any lumen already locked with heparin, lumen should be flushed with twice the indicated lumen volume using normal saline.Lumens should be flushed with normal saline between administrations of different infusates.After use, lumen should again be flushed with twice the indicated lumen volume using normal saline before reestablishing heparin lock.Based on the information provided, inspection of the returned product, and the results of the investigation, the cause could be established as an adverse event related to the patient condition.The appropriate personnel have been notified.Per the quality engineering risk assessment, no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SPECTRUM
Type of Device
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9763210
MDR Text Key181959927
Report Number1820334-2020-00480
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00827002478299
UDI-Public(01)00827002478299(17)211210(10)9510505
Combination Product (y/n)N
PMA/PMN Number
K081113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/10/2021
Device Catalogue NumberC-UTLM-701J-RSC-ABRM-HC-RD
Device Lot Number9510505
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2020
Initial Date Manufacturer Received 02/12/2020
Initial Date FDA Received02/27/2020
Supplement Dates Manufacturer Received04/09/2020
Supplement Dates FDA Received04/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age48 YR
Patient Weight92
-
-