Model Number 71441554
Device Problem Break (1069)
Patient Problem Full thickness (Third Degree) Burn (2696)
Event Date 02/10/2020
Event Type  malfunction  
Event Description
It was reported that during surgery, hi-flex impactor head size 3-8 broke while impacting the ps poly.All pieces were recovered from the patient.There was no harm to the patient or staff.No significant delay in the procedure because s&n back up was available.
 
Manufacturer Narrative
Additional information: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device fractured in half, rendering it inoperable.All pieces were returned.The device also had numerous nicks and gouges in the surface.The device was manufactured in 2006 and shows signs of extensive wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HIGH FLEX IMPCTR HD 3-8
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
Manufacturer Contact
sarah freestone
1450 brooks road
memphis, TN 38116
0447940038
MDR Report Key9774037
MDR Text Key181540806
Report Number1020279-2020-00747
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010497376
UDI-Public03596010497376
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441554
Device Catalogue Number71441554
Device Lot Number06HM20584
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/10/2020
Initial Date FDA Received03/02/2020
Supplement Dates Manufacturer Received02/10/2020
Supplement Dates FDA Received03/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-