Catalog Number UNK KNEE FEMORAL ADAPTOR
Device Problems Fracture (1260); Naturally Worn (2988)
Patient Problems Foreign Body Reaction (1868); Not Applicable (3189); No Code Available (3191)
Event Date 02/12/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Tc3 bolt and adaptor broke in patient.Outcome was a revison for tc3 femoral component.There was metal debris in the joint and the sleeve was well fixed.Surgeon removed the femoral component after 30 minutes of work.The 5 degree adaptor was still engaged in the sleeve which we were able to break the taper with slap hammer and five grip using the bolt.There have been multiple cases with the same problem which i have not been a part of.Surgeon was able to re-engage morse taper with an srom hinge implant.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.X-ray and implant photos were returned for review and confirmed the reported event of wear and fracture.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN KNEE FEMORAL ADAPTOR
Type of Device
UNK KNEE FEMORAL ADAPTOR
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9775005
MDR Text Key181737341
Report Number1818910-2020-06490
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK KNEE FEMORAL ADAPTOR
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/12/2020
Initial Date FDA Received03/02/2020
Supplement Dates Manufacturer Received04/06/2020
Supplement Dates FDA Received04/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PFC*SIGMA TC3 FEM RT SZ5; UNKNOWN KNEE FEMORAL ADAPTOR; UNKNOWN KNEE FEMORAL ADAPTOR BOLT; UNKNOWN KNEE FEMORAL AUGMENT; UNKNOWN KNEE FEMORAL AUGMENT; UNKNOWN KNEE FEMORAL SLEEVE; UNKNOWN KNEE TIBIAL INSERT
Patient Outcome(s) Required Intervention;
Patient Age64 YR
-
-