Reportable based on device analysis completed on 26feb2020.It was reported that a catheter kink occurred.A 85% stenosed target lesion was located at mildly calcified proximal left anterior descending artery.A 10/2.50 flextome cutting balloon was selected for use.During procedure, it was noted that the delivery shaft of the cutting balloon was kinked during advancing.The device was simply removed without any intervention and no fragments left inside patient body.However, device analysis revealed a complete break of the hypotube.The break was located at approximately 505mm distal of the strain relief.It was further reported that there was a complete break of the hypotube during the procedure.Subsequently, the device was removed successfully.The procedure was completed with another of the same device.No complications reported and patient was stable post procedure.
|