• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY MAILLEFER START-X TIP; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY MAILLEFER START-X TIP; SCALER, ULTRASONIC Back to Search Results
Catalog Number A066000000300
Device Problem Break (1069)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a start-x tip broke during use; no injury resulted.
 
Manufacturer Narrative
Returned start-x tip satelec insert 3 is broken in the active part.No material defect was found during analysis of the rupture pattern.The device shows no mark of excessive usury.Nothing unusual to report was found during dhr review (batch #1559965).The customer said having set his ultrasonic device at a power of 70.According to the dfu, the recommended power setting for start-x insert 3 connected to a satelec p5 newtron must be between 10 and 18.In all likelihood, the ultrasonic device used by the customer is not of this model, so we cannot attest that the selected setting complies maillefer recommendations.The root causes of the breakage are not identified.We will track this kind of event and monitor the trend.Note: the covid-19 pandemic caused a disruption in normal business activities, resulting in late submission of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
START-X TIP
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ  1338
MDR Report Key9776034
MDR Text Key184850012
Report Number8031010-2020-00028
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial,Followup
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA066000000300
Device Lot Number1559965
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2020
Initial Date Manufacturer Received 01/30/2020
Initial Date FDA Received03/02/2020
Supplement Dates Manufacturer Received03/17/2020
Supplement Dates FDA Received09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-