MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
|
Back to Search Results |
|
Model Number BI70000027230 |
Device Problems
Overheating of Device (1437); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.Foreign country: (b)(6).The manufacturer representative went to the site to test the imaging system.The reported issue was not confirmed.The cable connection on tube and thermometer were checked for loose contact.Possible contact problem on the molex connector (pull-off-protection not latched on one side).The cable connection checked on the atp board and generator for loose contact, no issue found.System checkout performed.The system was working as intended.) the manufacture date was not available at the time of reporting.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used in a sacroiliac and thoracolumbar procedure.It was reported that it was not possible to do x-rays.The pendant showed a message: "generator overheat - please wait".Imaging was aborted.No impact on patient outcome.No further information received.
|
|
Manufacturer Narrative
|
H2) additional information: event date updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3) a software analysis was initiated as part of the investigation.Log analysis found that the software of the imaging system functioned as designed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|