During the inguinal hernia repair with the uhi-4, the excessive pressure alarm of the subject device sounded.The user continued and completed the procedure by using the same uhi-4 with adjustment.After the procedure, the user found the subcutaneous emphysema on the patient and transferred the patient to the icu.Thereafter, the patient recovered and was discharged two days later.
|
The subject uhi-4 was returned to olympus medical systems corp.(omsc) for evaluation.During the evaluation by omsc, the reported phenomenon of the subject device, excessive pressure alarm, was not duplicated.Also omsc confirmed the subject device and found following.There was no abnormality on the outside and inside of the subject device.The function of the insufflation was inspected and there was no abnormality.Omsc reviewed the manufacturing history of the subject device and confirmed no irregularity.Since the reported phenomenon was not reproduced, the exact cause could not be conclusively determined.The instruction manual of the subject device states the cause of the excessive pressure alarm as follows, and the corresponding method in case of an abnormality.Additional gas insufflation from other equipment.The subject device setting.The patient's condition.
|