ST. JUDE MEDICAL PUERTO RICO, B.V. (CD) SJM MASTERS SERIES MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23AJ-501
Device Problem Material Separation (1562)
Patient Problem Mitral Insufficiency (1963)
Event Date 02/21/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.If returned, investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020 a sjm masters series mechanical heart valve was chosen for the procedure.The physician reported during the procedure, the device was implanted and the user noticed the leaflet detached.Additional information has been requested.2/24 pm.
 
Manufacturer Narrative
Additional information h10, h3 and h6 ; the reported event of the leaflets fractured and dislodged during rotation was confirmed.One leaflet had dislodged and fractured into multiple pieces, part of which was not returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the fractured leaflet.The cause of the reported event could not be conclusively determined, however it is consistent with damage caused by some external force applied to the leaflet and orifice, which overstressed the carbon material.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, B.V. (CD)
lot 21
caguas west industrial park
caguas, puerto rico 00726 -099
MDR Report Key9785874
MDR Text Key193266307
Report Number3003681312-2020-00006
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734021111
UDI-Public05414734021111
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/29/2024
Device Model Number23AJ-501
Device Catalogue Number23AJ-501
Device Lot Number7023832
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2020
Initial Date FDA Received03/04/2020
Supplement Dates Manufacturer Received03/19/2020
Supplement Dates FDA Received03/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-