• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC*CALIBRATED PAT CUT GDE; KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC*CALIBRATED PAT CUT GDE; KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS Back to Search Results
Model Number 86-5034
Device Problems Incorrect Measurement (1383); Device Slipped (1584)
Patient Problem Not Applicable (3189)
Event Date 02/17/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the black knob on the top part of the patella cutting guide spins when the saw is engaged in the cutting guide, resulting in an inaccurate cut.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned device found the device components to function as intended.The investigation found no evidence at this time of product malfunction or product error and the need for corrective action was not established.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9792792
MDR Text Key184534187
Report Number1818910-2020-06910
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295216599
UDI-Public10603295216599
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number86-5034
Device Catalogue Number865034
Device Lot NumberSO2043110
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2020
Initial Date Manufacturer Received 02/17/2020
Initial Date FDA Received03/05/2020
Supplement Dates Manufacturer Received03/04/2020
03/17/2020
Supplement Dates FDA Received03/13/2020
03/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age66 YR
Patient Weight88
-
-