Spacelabs was not able to collect the alarm logs from the central station due to the customer¿s delay in reporting the issue.The customer did provide a photo of the data in clinical access showing an extended run of ventricular tachycardia occurring from 8:20:00-8:20:10 on (b)(6) 2020.There was no vtach alarm recorder in the any logs for the waveform shown in the photo and it is unlikely that the condition would have been detected given the signal noise condition present in the photo.Spacelabs service technicians have tested the monitor that was used at the time of the alleged event and the device passed all functional testing.Due to the lack of data, further investigation is not possible.This report is complete and this particular issue is considered to be closed.
|