Model Number PSEE60A |
Device Problem
Failure to Advance (2524)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch # t5e557.Investigation summary: the analysis found that one psee60a device was returned with no apparent damage and with a gst60g cartridge reload loaded on the device.The cartridge was received partially fired 1/16 with eight double drivers out of position, making the reload non-functional.In addition, the cartridge pan was noted to be partially dislodged from left proximal side.The device was tested for functionality in the straight position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/batch.No conclusion could be reached on what may have caused the cartridge pan to dislodge.In addition, it should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure that the device meets the required specifications prior to shipment.
|
|
Event Description
|
It was reported that during a colon resection, when activating the stapler, the blade did not advance and the battery was discharged.A new stapler was used to complete the procedure.There were no patient consequences.No additional information is available at this time.
|
|