• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. POWERED 60 ECHELON +, 340MM SHAFT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. POWERED 60 ECHELON +, 340MM SHAFT; STAPLE, IMPLANTABLE Back to Search Results
Model Number PSEE60A
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/08/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t5e557.Investigation summary: the analysis found that one psee60a device was returned with no apparent damage and with a gst60g cartridge reload loaded on the device.The cartridge was received partially fired 1/16 with eight double drivers out of position, making the reload non-functional.In addition, the cartridge pan was noted to be partially dislodged from left proximal side.The device was tested for functionality in the straight position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/batch.No conclusion could be reached on what may have caused the cartridge pan to dislodge.In addition, it should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure that the device meets the required specifications prior to shipment.
 
Event Description
It was reported that during a colon resection, when activating the stapler, the blade did not advance and the battery was discharged.A new stapler was used to complete the procedure.There were no patient consequences.No additional information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERED 60 ECHELON +, 340MM SHAFT
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key9808664
MDR Text Key200622212
Report Number3005075853-2020-01518
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036014607
UDI-Public10705036014607
Combination Product (y/n)N
PMA/PMN Number
K110385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model NumberPSEE60A
Device Catalogue NumberPSEE60A
Device Lot NumberT94V45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2020
Initial Date Manufacturer Received 02/12/2020
Initial Date FDA Received03/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-