BOSTON SCIENTIFIC CORPORATION SOLYX SIS SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, FEMA Back to Search Results
Model Number M0068507000
Device Problems Detachment of Device or Device Component (2907); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).An examination of the returned solyx sis system device revealed that there was evidence of tearing on the mesh.Both carriers were detached from the mesh and were not returned.The shaft of the delivery device was bent; however, the deployment mechanism could still be extended.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.Based on the condition of the returned device, it is likely that the user experienced difficulty while advancing the device resulting in stretching the end of the mesh and the tearing it away from the carrier.Therefore, the investigation concluded that the most probable cause for this complaint is adverse event related to procedure, which indicates that the adverse event occurred during the procedure and the device had no influence on event.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
 
Event Description
It was reported to boston scientific corporation that a solyx sis system device was used during a solyx procedure performed on (b)(6) 2019.According to the complainant, while implanting the right part of the mesh during the procedure, the mesh carrier detached from the mesh and was removed from the patient by forceps.A photo of the mesh taken during the procedure revealed that the mesh appeared frayed.The procedure was completed with another solyx sis system device.There were no patient complications as a result of the event.The patient's condition after the procedure was reported to be stable.This event has been deemed reportable based on the investigation results: the shaft was bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLYX SIS SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, FEMA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key9808713
MDR Text Key190829869
Report Number3005099803-2020-00464
Device Sequence Number1
Product Code PAH
UDI-Device Identifier08714729774044
UDI-Public08714729774044
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K081275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/27/2022
Device Model NumberM0068507000
Device Catalogue Number850-700
Device Lot Number0023860804
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/10/2020
Initial Date FDA Received03/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1