• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX58OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX58OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS Back to Search Results
Model Number 1218-87-358
Device Problem Naturally Worn (2988)
Patient Problems Cyst(s) (1800); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Pocket Erosion (2013); Tissue Damage (2104); No Code Available (3191)
Event Date 04/22/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Pinnacle claim submission form and medical records received.After review of medical records, the patient was revised to address a failed left total hip arthroplasty with pain and signs of metallosis.Operative findings include severe reactive metallosis, soft tissue damage, erosion, significant amount of wear on the surface of the head and black metallic appearing rust over the trunnion.Samples taken from a cyst and pseudotumor were sent to the laboratory for analysis.Preoperatively, the patient was slightly longer on the right.Doi: (b)(6) 2004; dor: (b)(6) 2016 (right hip).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE MTL INS NEUT36IDX58OD
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9813981
MDR Text Key188470632
Report Number1818910-2020-07407
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K073504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1218-87-358
Device Catalogue Number121887358
Device Lot Number1198799
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/21/2020
Initial Date FDA Received03/10/2020
Supplement Dates Manufacturer Received04/14/2020
Supplement Dates FDA Received04/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-