Catalog Number 80395 |
Device Problems
Restricted Flow rate (1248); Difficult to Insert (1316); Temperature Problem (3022)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
While using a 30k fsi-sli-10s insert, it did not dispense water and the insert was getting warm; no injury resulted.
|
|
Manufacturer Narrative
|
Based on the evaluation on insert received, the insert meets temperature specification.The insert was also tested for water flow and passed water flow specification.The stack of the insert was found bent.The insert is beyond useful life.Note: the covid-19 pandemic caused a disruption in normal business activities, resulting in late submission of this report.
|
|