• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-04
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/25/2020
Event Type  Injury  
Event Description
The recipient is reportedly experiencing a potential device failure.Revision surgery is scheduled.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.The external visual inspection revealed silicone slices on the top cover prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.Advanced electrical testing performed on the device showed lower than typical range on some of the electrodes.It is the opinion of advanced bionics explant review board that some of these electrode test results should be considered as failures.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was initiated.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key9825883
MDR Text Key183250936
Report Number3006556115-2020-00108
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)170124(17)200131
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2020
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/12/2020
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
-
-