• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number N/A
Device Problem Wireless Communication Problem (3283)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/22/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was determined that the signal loss was related to the mobile application.It was indicated that the operating system for the smart device was not a supported system, which is off-label usage of the device.A review of the share logs was performed and signal loss was found within the investigation window.The allegation was confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Event Description
Subsequent to the initial mdr, a correction is required.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key9827017
MDR Text Key183505916
Report Number3004753838-2020-024607
Device Sequence Number1
Product Code QDK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K182041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberSW11586
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/23/2020
Initial Date FDA Received03/12/2020
Supplement Dates Manufacturer Received12/07/2021
Supplement Dates FDA Received12/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age30 YR
Patient SexMale
-
-