After review of medical record, patient was revised to address failed metal on metal left total hip arthroplasty.Revision notes stated that the posterior scar tissue was resected, there was minimal metallosis noted.Intraoperative x-rays confirmed a vertical and anteverted cup position.Push pull less than 1cm of leg length measurement.Patient had elevated metal ion and pain.Doi: (b)(6) 2008; dor: (b)(6) 2018; (left hip).
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|