Model Number 518-062
Device Problem Mechanical Jam (2983)
Patient Problems Iatrogenic Source (2498); Device Embedded In Tissue or Plaque (3165)
Event Date 02/20/2020
Event Type  Injury  
Event Description
A lead extraction case commenced to remove a non-functional right ventricle (rv) lead.A right atrial (ra) lead was also present in the patient but was not targeted for extraction.During preparation to removed the lead, a spectranetics lead locking device (lld) was placed inside the rv lead as a traction platform.Spectranetics tightrail rotating dilator sheath was in use in the innominate region and a glidelight laser sheath was in use in the subclavian region to aid in extraction.The physician then chose to use a cook medical byrd sheath to release lead on lead binding in the superior vena cava (svc) region.At the time the physician was using the cook byrd sheath, the patient's blood pressure dropped.Rescue efforts began immediately, including rescue device and sternotomy.Perforations were identified in the superior vena cava (svc) and right atrial (ra) junction.The perforations were repaired successfully, and the patient survived the procedure.During the repair, the physician attempted to unlock and remove the lld.When the attempt to unlock the lld was unsuccessful, the lld and rv lead was cut, capped, and left within the patient's body.This report is being submitted to capture this lld device.No spectranetics devices were reported to have caused or contributed to the patient injuries mentioned above.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
jessica dominguez
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key9843743
MDR Text Key184075656
Report Number1721279-2020-00050
Device Sequence Number1
Product Code DRB
UDI-Device Identifier00813132023076
UDI-Public(01)00813132023076(17)210830(10)FLP19H28A
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/30/2021
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberFLP19H28A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 02/20/2020
Initial Date FDA Received03/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age33 YR
Patient Weight157
-
-