• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Pressure Problem (3012)
Patient Problems Retinal Detachment (2047); Retinal Tear (2050); Collapse (2416)
Event Date 02/21/2020
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgeon reported there was a loss of infusion pressure during a procedure.The patient experienced an eye collapse and retinal tear.Additional information has been requested and received.Additional information has been received from the surgeon which indicated the patient had proliferative diabetic retinopathy (pdr) with retinal traction, and was under going a vitrectomy procedure with laser treatment for this issue.The surgery was going as planned with the expected issues of mild corneal haze, blood stuck to back surface of lens, epi retinal membrane (erm) and retinal tractions.After removal of the tractions and peeling of the erm the retina was observed to be completely flat with no obvious breaks.The patient felt some discomfort during the indentation process of the procedure so additional anesthesia was administered subtenons.The laser portion of the procedure was performed under air.The surgeon went back to fluid/air exchange (fax) and the eye suddenly collapsed.A break in the retina was observed 180 degrees opposite the inferotemporal infusion line.A small air bubble had escaped subretinally as well.A rrhegmategenous retinal detachment was noted.The break was sealed using cryotherapy and gas.
 
Manufacturer Narrative
Additional information has been provided in h.6.And h.10.The customer did not request service for the system.The system serial number (s/n) was not provided and could not be determined based on the information provided.Therefore, manufacturing information could not be obtained.With no additional, related information provided, the customer reported event could not be confirmed.The root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The company service representative (ar) examined the system on and could not replicate the reported issue.The system was then tested and met all product specifications.Based on assessment, the product met specifications at the time of release.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key9853105
MDR Text Key184069788
Report Number2028159-2020-00249
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 06/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/21/2020
Initial Date FDA Received03/19/2020
Supplement Dates Manufacturer Received04/29/2020
06/03/2020
Supplement Dates FDA Received05/04/2020
06/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
-
-