This report has been identified as b.Braun melsungen ag internal report # (b)(4).Device history record (dhr): reviewed the dhr for the batch no.18d16g8f01, no abnormalities observed during the in-process and final control inspection.Visual inspection: the perifix catheter is broken off approx.83 mm away from the catheter end.The shorn off part with the catheter beginning was not handed over by the customer.Summary and assessment: such damages may occur when the catheter will be withdrawn against the cannula bevel and thereby shear off.We exclude a manufacturing fault since the catheters were taken to a 100% examination and therefore no mechanical damages or manufacturing faults are allowed.Please see instructions for use: "never pull the catheter through the needle as it may otherwise shear off." therefore we assume a fault during the application process and consider the complaint as not confirmed.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|