Catalog Number 402439
Device Problems Loose or Intermittent Connection (1371); Device Dislodged or Dislocated (2923); Migration (4003)
Patient Problem Failure of Implant (1924)
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as dislocation, loosening and migration.The actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation, loosening, and migration.There were no findings during this evaluation that indicate that the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Reported incident: the products of djo were associated products in a complaint regarding dislocation, loosening, and migration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE CEMENT
Type of Device
COBALT MV WITH GENTAMICIN (AKA COBALT G-MV) BONE CEMENT
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key9854015
MDR Text Key189885693
Report Number1644409-2020-00199
Device Sequence Number1
Product Code LOD
UDI-Device Identifier00888912338387
UDI-Public(01)00888912338387
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092150
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2012
Device Catalogue Number402439
Device Lot Number632240
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/19/2020
Initial Date FDA Received03/19/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-