Model Number EP003994S
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/13/2015
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the transseptal needle, ep003994s with lot number 209002496, was returned and analyzed.There were no anomalies found.The reported skiving issue was not confirmed; however, it is plausible that skiving occurred as there is a compatibility issue with the needle and a competitor sheath.The cause of the issue was not established.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that following a successful puncture with the transseptal needle the physician retracted the needle from the sheath and attempted to advance a wire through the sheath.The wire would not pass through the sheath; there appeared to be a blockage.The sheath was exchanged and no patient complications have been reported as a result of this event.Additional information reported that the dilator was scratched.Plastic parcels were found after removal of the sheath from the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH¿ NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9857945
MDR Text Key193661460
Report Number9612164-2020-01239
Device Sequence Number1
Product Code DRC
UDI-Device Identifier00613994663146
UDI-Public00613994663146
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/26/2016
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number209002496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2015
Initial Date Manufacturer Received 07/23/2015
Initial Date FDA Received03/19/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/04/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-